Laserfiche WebLink
This document is the Intellectual Property of SGS AXYS Analytical Services Ltd. and contains Proprietary and Confidential Business <br />Information. It may not be reproduced or distributed without written permission of the owner. © SGS AXYS Analytical Services Ltd, 2020. <br />SGS AXYS Analytical Services Ltd, 2045 Mills Road W., Sidney, BC, Canada, V8L 5X2. Tel. (250) 655-5800, fax (250) 655-5811. <br />This is not a controlled document; it is subject to change without notification. <br /> <br />SGS AXYS Analytical Services Ltd. <br /> <br />MSU-110 Rev 14, 28-May-2020 Summary of MLA-110 Rev 02 Ver 07 Page 16 of 23 <br />SGS AXYS confidential, not to be distributed without written permission <br />2 These surrogates used only to quantify the analogous native compounds. Formal surrogate recovery limits are under <br />review. <br />3 For DoD: Surrogate recoveries are determined by comparing surrogate peak area in samples to surrogate’s area <br />measured in the day’s initial cal/ver (or I-CAL if run that day). <br />QC Specification Table: Other Parameters <br />QC Parameter Specification <br />MS Acquisition Rate <br /> <br />Minimum acquisition rate for every native analyte and labeled compound <br />peak: At least 10 data points per peak. <br />Instrument Sensitivity <br /> <br />Daily, S:N ≥ 10:1 for the primary transition product ion for all analytes for the <br />lowest calibration standard (CAL B) and S:N ≥ 3:1 for the secondary ion. <br />Additional requirement for DoD: Prior to sample analysis and then at least <br />once every 12 hours, verify sensitivity at the LOQ level by running CAL C <br />The found analyte concentrations must be within ±30% of the true values. <br />Mass Calibration Instrument must have a valid mass calibration following the manufacturer <br />specified procedure prior to any sample analysis. The mass calibration is <br />updated on an as-needed basis (e.g. QC failures, ion masses fall outside of <br />the ±0.5 amu of the true value, major instrument maintenance, or if the <br />instrument is moved.) <br />Refer to SIN-033. The entire range (bracketing all the masses of the target <br />analytes) must be mass calibrated. The maximal allowed residual error is ≤ <br />0.1 Da for each mode with no more than two calibration points missed. <br />Mass Calibration <br />Verification <br />Mass calibration is verified after each mass calibration, prior to initial ICAL. <br />Mass calibration must be verified to be ±0.5 amu of true value by acquiring <br />a full scan continuum mass spectrum of a PFAS stock standard OR by <br />following the instrument manufacturer’s instructions for performing a mass <br />calibration verification and using the instrument manufacturer’s <br />recommended standards as long as these standards cover the mass range <br />of the PFAS ions of interest. <br />Initial Calibration (I-CAL) <br /> <br />Run initially, and as required to maintain calibration verification and <br />instrument sensitivity. CAL A is a sensitivity standard S:N ≥ 3:1 for the <br />primary transition product ion for all analytes. CAL B is the default lowest <br />calibration standard. (CAL A is not included in determination of the RRF). <br />Quantification is achieved by the constant RRF method. The I-CAL <br />specifications for the RRF fit are <20% RSD of mean RRFs and 70-130 % <br />recovery of analytes at each concentration level down to C level. The <br />recovery specification for the B CAL is 50-200%. The surrogate recovery <br />specifications are 50-150% recovery at each concentration level. <br />Peak Asymmetry: 0.8-1.5 for PFBA and PFPeA measured in CAL E (mid <br />cal point) at 10% of the peak height. If this is not achieved, perform <br />instrument maintenance and re-run I-CAL. <br />www.axysanalytical.com <br />Page 21 of 94